Skip to content

Phasix ESD Andover semiconductor test laboratory

Hi Gain trading as Phasix ESDAndover, Hampshire

Phasix ESD’s Andover laboratory tests packaged electronic components and semiconductor devices. Current UKAS 4670 issue 012 covers HBM, CDM and MM electrostatic-discharge methods at the main address only, with separate ranges, fixture conditions and method exclusions. The owner also offers latch-up and TLP testing outside that schedule. Agree the device, pin groups, exact standard and reporting requirements with the provider.

Reviewed by Arc · About this review

Key facts

Exact test site
Unit 17, Focus 303 Business Centre, Focus Way, Andover, Hampshire SP10 5NY, United Kingdom. The owner and schedule identify this address. The UKAS result calls the same site Business Park. Testing is scheduled at the main address only.
Current official schedule
UKAS laboratory 4670 is accredited to ISO/IEC 17025:2017 for its printed methods. Issue 012 is dated 10 November 2025 and both complete pages were inspected on 1 October 2026. HBM, CDM and MM ranges and exclusions remain separate.
Scope distinction
Latch-up and TLP are owner-offered semiconductor services outside the retained UKAS schedule. The robotic HBM 100 mm × 100 mm package claim is owner advertising, without an accredited package-size envelope inferred.
Published enquiry contact
Use +44 1264 361622 or the schedule’s carl.simon@phasix.co.uk. The schedule names Carl V R Simon as its published contact, without a present job-role claim. The owner contact page corroborates the phone and test address but leaves its visible email value empty.

Testing capabilities

Scheduled human body model ESD testing

For packaged electronic components and semiconductor devices, HBM reaches 8 kV. The Mk2 system uses fixtures up to 768 pins, while the schedule separately states no pin or fixture limitation for HPPI ATS 8000G. JS-001 editions through 2024 exclude pin reduction sampling methods. MIL-STD-883G 3015.7, 883H 3015.8 and 883J 3015.9 each exclude section 2.3.2. The printed method and system conditions control the work.

Scheduled charged device model ESD testing

The CDM row for packaged electronic components and semiconductor devices reaches 2 kV, independently of the HBM range. It lists AEC-Q100-011 revisions B:2003, C1:2013 and D:2019, AEC-Q101-005:2005, JESD22-C101 editions C through F and JS-002 editions 2014, 2018, 2022 and 2025. Confirm the exact edition and device arrangement.

Scheduled machine model ESD testing

The separate MM row reaches 2 kV and uses fixtures up to 768 pins. It lists EIA/JESD22-A115-A:1997 and -C:2010, AEC-Q100-003 revision E:2003 and AEC-Q101-002 revision A:2005. These are the printed schedule references, without later editions or other-system capacities inferred.

Owner-offered latch-up and TLP testing

The owner offers latch-up testing and transmission-line-pulse characterisation for semiconductor devices. Its services page lists JESD78 variants for latch-up and ESD TR5.5-01-08 / ANSI/ESD STM 5.5.1-2016 for TLP. These offers are outside the current UKAS 4670 schedule. Agree the method, fixture, item and report status directly.

Equipment and published limits

Published equipment limits and conditions
SpecificationPublished valueConditions
Scheduled HBM voltage and fixture conditionsUp to 8 kV, Mk2 fixtures up to 768 pinsHBM row only. The schedule separately states no pin or fixture limitation for HPPI ATS 8000G. This does not establish an unrestricted package size, interchangeable tester capacity or calibration status.
Scheduled CDM voltageUp to 2 kVCDM row for packaged electronic components and semiconductor devices. Separate from HBM and MM methods, without combined operating limits inferred.
Scheduled MM voltage and fixturesUp to 2 kV, fixtures up to 768 pinsMM row only. The schedule lists Verifier V2/V3, Thermo KeyTek RCDM 3, Thermo Mk. 2 SE/768 and HPPI ATS 8000 G systems without assigning every capacity to every system.
Owner robotic HBM package-size statementDevices up to 100 mm × 100 mmOwner description of its robotic two-pin HBM tester. Package fit, fixtureless operation and pin compatibility require confirmation. The numerical package limit is not a current scheduled accreditation range.

Access and contact

Contact Phasix ESD on +44 1264 361622 or use carl.simon@phasix.co.uk from the current schedule. The owner discusses testing type, pin combinations, parameters and sample quantity, and offers test-plan agreement on request. Agree the device, package and fixture, stress method/edition, functional-test responsibilities, required report and programme before requesting work.

Facility contact

Address
Unit 17, Focus 303 Business Centre, Focus Way
Andover, Hampshire, SP10 5NY
United Kingdom

Standards and accreditation

UKAS 4670 exact main-address scope
Issue 012 dated 10 November 2025 covers HBM, CDM and MM at the Andover main address only. HBM includes JS-001-2024 with pin reduction sampling excluded and the printed MIL section 2.3.2 exclusions. CDM includes JS-002-2025. Exact scheduled editions and device/method/system conditions control accredited claims.
Owner-only methods and certification wording
Latch-up and TLP standards on the owner site are outside the retained schedule. ISO 9001 certification and the owner’s statement about future accreditation of latest standards do not extend current accredited scope. The robotic package-size claim and general turnaround advertising are separate from scheduled method limits.

What to confirm with the facility

  • Use the exact HBM, CDM or MM row and main test address. HBM section 2.3.2 and pin reduction sampling exclusions, Mk2 fixture conditions and the separate HPPI statement must remain attached to the requested method. The owner’s 100 mm × 100 mm package statement is not an accredited numerical range.
  • Owner latch-up and TLP offers are outside the current schedule. Confirm method, fixture compatibility, sample quantity, report status, functional testing and programme. Published five-day turnaround and future-accreditation statements are not an unconditional guide commitment.
  • Provide the device identity, package drawing, pin assignment, supply limits, intended standard, stress levels, sample quantity and report requirements so the provider can confirm suitability and cost.

Before requesting a test

Public information not available. Contact test facility provider for more information.

Frequently asked questions

Public information not available. Contact test facility provider for more information.

Public information not available. Contact test facility provider for more information.

Further reading

Public information not available. Contact test facility provider for more information.

Sources and review

Reviewed by Arc on 1 October 2026 in Europe/London using five fresh official sources. Both complete pages of UKAS 4670 issue 012 were visually inspected. The Andover main address, separate HBM/CDM/MM ranges, fixture conditions, method editions and exclusions are mapped. Owner latch-up/TLP methods and robotic HBM package-size advertising remain outside the current schedule. Business Centre in the owner/schedule versus Business Park in the register is preserved. The schedule contact email and named-contact wording are distinct from the owner’s empty visible email and Southend registered office. Original bodies, headers, retrieval metadata, titles, absent owner canonicals/dates, UKAS template dates, PDF extraction and full page renders are retained. Issue and PDF dates on 10 November 2025 are separate records from this review. Prior owner and schedule discovery bodies remain byte-identical and unchanged. Imported postcode coordinates are cleared.

Arc publishes this directory entry and does not operate the test facility. Suggest a correction to Arc with a supporting source.

  1. UKAS Phasix ESD schedule 4670, issue 012

    United Kingdom Accreditation Service · Reviewed

    Current schedule 4670 issue 012 dated 10 November 2025 is limited to the main Andover address. Both complete pages were visually inspected. HBM up to 8 kV has Mk2 fixture and HPPI ATS 8000G distinctions, section 2.3.2 exclusions for the printed MIL methods and a pin-reduction-sampling exclusion for the JS-001 list. CDM and MM separately reach 2 kV. JS-001-2024 and JS-002-2025 are listed. Latch-up and TLP are absent. PDF creation/modification at 14:20:05 UTC on 10 November 2025 is separate from the issue date and this review.

  2. UKAS Hi Gain trading as Phasix ESD laboratory 4670 result

    United Kingdom Accreditation Service · Reviewed

    The current UKAS result links laboratory 4670 to Hi Gain Ltd T/A Phasix ESD at Unit 17, Focus 303 Business Park, Focus Way, Andover SP10 5NY, United Kingdom. The schedule and owner contact page call the same site Focus 303 Business Centre, retained for the address. Directory-template publication on 30 June 2020 and modification on 11 November 2022 do not date this result.

  3. Phasix ESD official laboratory overview

    Hi Gain trading as Phasix ESD · Reviewed

    The owner describes semiconductor HBM, MM, CDM, latch-up and TLP services. Its robotic HBM package-size claim of 100 mm × 100 mm is kept as owner advertising, separately from the schedule’s fixture/pin scope. Claims about parts exceeding 2,000 pins, rapid results and guaranteed turnaround are not allocated as unconditional guide promises. ISO 9001 wording does not establish method accreditation. Original title is Phasix ESD, with no canonical or publication/modification metadata.

  4. Phasix ESD official contact details

    Hi Gain trading as Phasix ESD · Reviewed

    The owner contact page corroborates Hi Gain Ltd T/A Phasix ESD, Unit 17, Focus 303 Business Centre, Focus Way, Andover SP10 5NY and the phone. The Southend-on-Sea registered office is not the test site. The visible page has an empty email value, so the guide’s carl.simon@phasix.co.uk route comes from the current schedule. Original title is Contact Phasix ESD, without canonical or page-date metadata.

  5. Phasix ESD official testing services

    Hi Gain trading as Phasix ESD · Reviewed

    The owner offers device qualification and investigative ESD/latch-up work, customer discussion of pin combinations, parameters and samples, and test-plan agreement on request. Latch-up and TLP standards are owner offers outside the retained UKAS schedule. The prospective latest-standards accreditation note and ISO 9001 certification wording do not extend the current schedule. Report/media/turnaround and 48-hour testing statements are not unconditional guide commitments. Original title is Phasix ESD Services, without canonical or page-date metadata.